Channel A Frequency Response And Sensitivity Test, 10 Hz-20 Mhz - HP 5334B Service Manual

Universal counter
Hide thumbs Also See for 5334B:
Table of Contents

Advertisement

Equipment:
Function Generator
..............................................................................................
H P 332%
Procedure:
1. Set the function generator as follows:
.......................................................... ...................................................
Frequency
10 Hz
Amplitude
..............................................................................................................
15 mV rms
Function
.................................................................................................................
Sine Wave
2. Set the H P 5334B as follows:
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
FREQ B
SENS
.......................................................................................................................
ON
...........................................................................
CHAN B TRIGISENS control
fully CW
3. Connect the function generator signal to the HP 5334B Input B using a 50 Ohm feedthrough connector as
shown in Figure 4-3.
4.
Repeat the tests of paragraph 4-31 for Channel B and record the measurements on the Performance Test
Record Card, lines 7 through 12. Begin the tests at the verification of 10 Hz in paragraph 4-31, step 4.
Failure: If any of these tests fail, refer to Section V, Adjustments, paragraphs 5-17 and 5-19 as a first step in
troubleshooting.
e
4-30. CHANNEL A FREQUENCY RESPONSE AND SENSITIVITY TEST,
80 MHz-100 MHz
This test is for instruments that do
not
contain the Option 060 Rear Panel Inputs, i.e., instruments with Front In.
puts only.
Specification: Refer to Table
1-1, HP
Model 5334B Specifications, for Frequency Response and Sensitivity
specifications.
Description: The frequency measuring range of the Counter is tested at minimum sensitivity specifications and
two different frequency settings.
Frequencies and conditions tested:
80 MH7 and 100 MHz, dc coupled, 1 MR
80 MHz and 100 MHz, ac coupled, 1 MR
80 MHz and 100 MHz, ac coupled, 50R
HP 5334B
-
Service Manual
4-17