Equipment:
..............................................................................................
Function Generator
HP 3325A
Procedure:
1.
Set the function generator as follows:
...............................................................................................................
Frequency
5 MHz
..............................................................................................................
Amplitude
100 mV rms
Function .................................................................................................................Sine
Wave
2. Set the HP 5334B as follows:
Reinitialize the HP 5334B.
FUNCTION
...........................................................................................................
RATIO A/B
SENS
.......................................................................................................................
ON
A & B TRIGISENS controls
...............................................................................
fully CW
CHAN
A & B
...............................................................................................ON
3. Connect
HP
5334B TIME BASE signal to Input A and connect the function generator signal to Input B.
VERIFY: The Counter displays 2.000 000
+
000 001.
4. Record the Ratio
A/B
measurement on the Test Record Card, line 37.
Failure: If the instrument under test does not meet the test specification, consider performing the adjustments in
Section V of this manual as
a
first step in correcting the problem.