HP 5334B Service Manual page 17

Universal counter
Hide thumbs Also See for 5334B:
Table of Contents

Advertisement

4. Set the
H P
5334B as follows:
COM A
...................................................................................................................
OFF
CHAN A and B 500
............................................................................................
OFF
Observe: The
H P
5334B front panel displays 10.000 000 0 MHz
+
0.2
Hz.
Channel A trigger light flashing
and Channel B light not flashing.
5. Set the
H P
5334B as follows:
CHAN A
50i2
........................................................................................................
ON
Observe: The
H P
5334B front panel displays
all
dashes and Channel A trigger light stops flashing.
6. Set:
........................................................................................................
CHAN A 50 R
OFF
Observe: Condition prior to switching in 50R impedance.
7. Set:
CHAN A XI0 A'ITN
...........................................................................................
ON
Observe: The
H P
5334B front panel displays all dashes and Channel A trigger light stops flashing.
8. Set:
...........................................................................................
CHAN A XI0 ATTN
OFF
Observe: Condition prior to switching in XI0 attenuator.
9. Set:
100
kHz
FILTER A
..............................................................................................
ON
Observe: The
H P
5334B front panel displays all dashes and Channel A trigger light stops flashing.
10. Set:
..............................................................................................
100
kHz
FILTER A
OFF
Observe: Condition prior to switching in 100 kHz Filter.
1 1 . Set:
...................................................................................................................
COM A
ON
...........................................................................................................
FUNCTION
FREQ B
Connect the rear panel TIME BASE oscillator signal to the Channel B Input.
Observe: The
H P
5334B front panel stops updating and the trigger lights stop flashing.
12. Set:
...................................................................................................................
COM A
OFF
Observe: The
HP
5334B front panel displays 10.000 000 0 MHz
2
0.2 Hz. Channel B trigger light flashing
and Channel A trigger light not flashing.
13. Set:
CHAN B 50R
........................................................................................................
ON
H P
5334B
-
Service Manual
4-7

Advertisement

Table of Contents
loading

Table of Contents