4-35. PULSE WIDTH
A
TEST
Specification: Refer to Table
1-1,
HP Model 5334B Specifications, for Pulse Width
A
specifications.
Description: A pulse width is generated and then measured with the HP 5334B to verify the Counter's perfor-
mance.
I
Figure 4-9. Pulse Width
A
Test Setup
P
\
Equipment:
Signal Generator
...................................................................................................
HP 8660AIC
HP5334B
UNIVERSAL COUNTER
Procedure:
1. Set the signal generator as follows:
Frequency
...............................................................................................................
100 MHz
Amplitude
..............................................................................................................
200
m V
rms
Q
r r
J
HP8660A/C/D
SIGNAL GENERATOR
2. Set the HP 5334B as follows:
REF
OUT
Reinitialize the 5334B.
FUNCTION
...........................................................................................................
PULSE WIDTH A
'
HP 5334B
-
S e ~ c e Manual
4-26