Table of Contents

Advertisement

4.
Adjust AlCl (OSC ADJ) for a 5334B front panel display of 10.000 000 MHz
2
8 Hz.
5. Set the oscilloscope as follows:
Channel
A
V/Div.
...........................
0.5V
Time/Div
...........................................
0.05ps
Ch A Input Coupling
......................
5051
Main Triggering
..............................
INT
Internal Trigger
...............................
A
Vertical Display
...............................
A
Horizontal Display
..........................
MAIN
6. Connect the 5334B TIME BASE output to the oscilloscope channel A input.
7. Verify that the signal is greater than 1.4V p-p.
NOTE
The following procedure is for the Option 010 Oven Oscillator equipped
53348.
2
HP1715A
OR
EQUIVALENT
OSCILLOSCOPE
TSlB-L2
Figure 5-5. Option 010 Oscillator Adjustment Setup
Equipment:
Oscilloscope
.....................................
HP 1715A (or equivalent)
Ref. Freq. Std.
.................................
10 MHz
Reference: Figure 8-27
C
HP
5334B
- Service
Manual
5-8