6. Record the measurement on the Performance Test Record Card, line 14.
...............................................................................................................
7. Set the signal generator as follows:
Frequency
80 MHz
Amplitude
..............................................................................................................
35 mV rms
8. Set the HP 5334B as follows:
...........................................................................................................................
AC
ON
VERIFY: The Counter displays 80 MHz r 2 Hz.
9. Record the measurement on the Performance Test Record Card, line 15.
10. Set the signal generator as follows:
Frequency
.............................................................................................................
1 0 0 MHz
Amplitude
..............................................................................................................
35 mV rms
VERIFY: The Counter displays 100 MHz
+
2 Hz.
1 1 . Record the measurement on the Performance Test Record Card, line 16.
12. Set the signal generator as follows:
Frequency
...............................................................................................................
80 MHz
...................................
Amplitude
:
.........................................................................
35 mV rms
NOTE
Remove the 50R feedthrough connector.
13. Set the HP 5334B as follows:
..........................................................................................................................
50R
ON
VERIFY: The Counter displays 80 MHz
+
2
Hz.
14. Record the measurement on the Performance Test Record Card, line 17.
15. Set the signal generator as follows:
Frequency
...............................................................................................................
100 MHz
Amplitude
..............................................................................................................
35 mV rms
VERIFY: The Counter displays 100 MHz
+
2 Hz.
16. Record the measurement on the Performance Test Record Card, line 18.
Failure: If any of these tests fail, refer to Section V, Adjustments, paragraphs 5-16 and 5-18 as a first step in
troubleshooting.
HP 5334B
-
Service Manual
4- 19