Analog Output Range High
Set the engineering range value, in engineering units, at which maximum output is required.
Analog Output Range Low
Set the engineering range value, in engineering units, at which minimum output is required.
Analog Electrical Range High
Set the maximum current output value to correspond with the Analog Output Range High,
between 0.0 and 20.0 mA in 0.1 mA increments.
Analog Electrical Range Low
Set the minimum current output value to correspond with the Analog Output Range Low,
between 0.0 and 20.0 mA in 0.1 mA increments.
Return to top of page.
3 Configuration
IM/C1300 Rev. J