ABB C1300 User Manual page 77

Advanced circular chart recorder
Hide thumbs Also See for C1300:
Table of Contents

Advertisement

C1300
Advanced circular chart recorder
°C
Type
Range Low
B
–18
E
–100
J
–100
K
–100
N
–200
R
–18
S
–18
T
–250
PT100
–200
Table 8.2 Thermocouple Performance
2-Wire Transmitter Power Supplies
Number
Voltage
Drive
Isolation
Analog Outputs
Type
Accuracy
Maximum load
Dielectric
Relay Outputs
Type
Rating (with non-inductive load) 5A at 115/230V AC
Digital Inputs
Type
Minimum pulse
Dielectric
Digital Outputs
Type
Rating
Dielectric
Serial Communications
Connections
Protocol
IM/C1300 Rev. J
Range High
Range Low
1800
0
900
–140
900
–140
1300
–140
1300
–325
1700
0
1700
0
300
–400
600
–325
1 per channel
24V DC nominal
Up to 25mA
500V DC channel-to-channel
4 to 20mA
±0.1%
750W
500V DC
SPDT
TTL or volt-free
250ms
500V DC between modules,
no isolation within module
5V TTL
5mA per output
500V DC between modules,
no isolation within module
RS485, 4-wire
Modbus RTU
°F
Range High
3270
±2.0°C (above 200°C) (3.6°F [above 434°F])
1650
±0.5°C (0.9°F)
1650
±0.5°C (0.9°F)
2350
±0.5°C (0.9°F)
2350
±0.5°C (0.9°F)
3000
±1.0°C (above 300°) (1.8°F [above 572°F])
3000
±1.0°C (above 200°C) (1.8°F [above 434°F])
550
±0.5°C (0.9°F)
1100
±0.5°C (0.9°F)
8 Specification
Accuracy (excluding CJC)
75

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents