Measurement Algorithms - HP 16534A User Reference

1-gsa/s and 2-gsa/s digitizing oscilloscope
Hide thumbs Also See for 16534A:
Table of Contents

Advertisement

The Auto-Measure Menu

Measurement Algorithms

Measurement Algorithms
Frequency (Freq)
The frequency of the first complete cycle displayed is measured using the
50% levels.
If the first edge on the display is rising, then
Freq =
t
rising edge 2
If the first edge on the display is falling, then
1
Freq =
− t
t
falling edge 2
Period
The period is measured at the 50% voltage level of the waveform.
If the first edge on the display is rising, then
Period = t
rising edge 2
If the first edge on the display is falling, then
Period = t
falling edge 2
Peak-to-Peak Voltage (Vp_p)
The maximum and minimum voltages for the selected source are measured:
− V
Vp_p = V
maximum
where V
and V
maximum
minimum
present on the selected source.
8-8
1
.
− t
rising edge 1
.
falling edge 1
− t
.
rising edge 1
− t
.
falling edge 1
,
minimum
are the maximum and minimum voltages

Advertisement

Table of Contents
loading

This manual is also suitable for:

16533a

Table of Contents