Chapter 3: Testing Performance
To test voltage measurement accuracy
To test voltage measurement accuracy
This test verifies the voltage measurement accuracy of the instrument. The measurement is
made in a way that offset errors are not a factor.
Specification
Dual Cursor
+(1.25% of full scale + 0.032 x V/div)
Single Cursor
+{1.25% of full scale + offset accuracy + (0.016 x V/div)]
Equipment Required
Equipment
Critical Specifications
Power Supply
7 mV to 30 Vdc, 0.1 mV resolution
Digital Multimeter
Better than 0.1% accuracy
Cables (2)
BNC
Adapters (2)
BNC (f} to banana (m}
Adapters (2)
BNC tee (m)(f)(f)
Blocking capacitor
0.18 LF
Shorting cap
BNC
Procedure
Recommended
Model/Part
HP 61144
HP 34401A
HP 10503A
HP 1251-2277
HP 1250-0781
HP 10240B
HP 1250-0774
A power supply provides a reference voltage to check measurement accuracy. The supply is
monitored for accuracy (especially at low voltages). A dc blocking capacitor is used to filter
noise on the input voltage to the oscilloscope,
1 Connect the equipment.
a Use a BNC-to-banana adapter to connect a BNC cable to the power supply.
b Connect a BNC tee to the other end of the cable and use a BNC-to-banana adapter to
connect the tee to the DVM
e Connect another BNC cable to the tee at the DVM and connect a BNC tee to the cable.
d Connect the blocking capacitor to the BNC tee and connect the BNC shorting cap to
the blocking capacitor.
FROW
POWER SUPPLY
TO OM
aNC-To-
i AN
BANANA
OLN
O
ADAPTERS
SNC CABLE
a i
TT
gNC TEL
uc
10
CHANNEL
ENSUE
BLOCKING
BNC SHORT
a
CAPACITOR
mi)
TEE
[A
{NETO
BNC CABLE
Re'
{
|
Voltage Measurement Accuracy Equipment Setup
S45a2e11