To test trigger sensitivity
To test trigger sensitivity
This test checks channel and external triggers for sensitivity at rated bandwidth.
Specification
dc to 100 MHz
100 MHz to 500MHz
Internal
0.5 div
1.0 div
External
0.0225 x signal range
0.045 x signal range
Auxiliary
dc to 50 MHz, 250 mVp.p
Equipment Required
Equipment
Critical Specifications
Recommended
Model/Part
Signal Generator
100 and 500 MHz, 30-80 mVrms output
HP 8656B
Power Splitter
outputs differ by <0.15 dB
HP 11667A
Termination
BNC feedthrough, 50 Q
HP 10100C
Cable
Type N (m) 24 inch
HP 11500B
Cable
50 Q BNC 36 inch
HP 10503A
Adapter
Type N (f} to BNC (m)
HP 1250-0077
Adapter
Type N {m) to BNC {m)
HP 1250-0082
Adapter
Type N (m) to BNC {f)
HP 1250-0780
Adapter
BNC tee (mi(f(f)
HP 1250-0781
Internal Trigger Test
Perform this test on all vertical channels.
Press Recall ,thenpress
Cir
to set the oscilloscope to default conditions.
Set the following parameters
Menu (hardkey)
Function/Parameter
Setting
Vertical
1
on
{channel key)
others
off
{all}
(sensitivity)
200 mV/div
{input R)
50 0C
Horizontal
{time/div)
5 ns/div
(Setup)
(mode)
repetitive
Display
(mode)
avg
# of avg
16
(graticule)
grid
With an N cable and N-to-BNC adapter, connect the signal generator to the channel 1
input.
3-15