HP 54520 Series Service Manual page 87

Hide thumbs Also See for 54520 Series:
Table of Contents

Advertisement

PERFORM
SIGNAL
TESTS
ON ALL
CHANNELS.
NOTE
WHICH
CHANNELS
FAIL.
To troubleshoot the instrument
CHART
6A
REPLACE
FAULTY
ATTENUATOR.
PUT GOOD
(SEE
ATTENUATOR
TROUBLE—
SHOOTING. )
SWAP FATLING
CHANNEL
ATTENUATOR
WITH A GOOD ONE
54504808
BE SURE DEFAULT CALS ARE
LOADED
(CHART
1)
PERFORM GLICK TEST ON ALL
CHANNELS.
NOTE WHICH
CHANNELS
FAIL.
(SEE
ATTENUATOR
TROUBLESHOOTING)
00
ANY CHANNELS
FAIL?
SWAP FAILING
CHANNEL
ATTENUATOR
WITH A GOOD ONE
TEST ON THE
TWO SWAPPED
CHANNELS:
MEASURE
INPUT
IMPEDANCE
OF EACH
CHANNEL .
{SEE INPUT
RESISTANCE
PERFORMANCE
TEST
SECTION
3)
5@ OHM
PERFORM
SIGNAL
TEST ON THE
TWO SWAPPED
CHANNELS:
PERFORM CLICK
ATTENUATOR
INTO
ORIGINAL
CHANNEL
REPLACE
MAIN
ASSEMBLY
GO
TO CHART
1
SAME ATTENUATOR
NO
REPLACE
FAULTY
ATTENUATOR.
PUT
GOOD
ATTENUATOR
INTO
ITS
ORIGINAL
CHANNEL
AND
1M OHM
?
REPLACE
ATTENUATOR
MEASURE
EXT
TRIG. IMPEDANCE
{CHAN
MENU-EXT}
5@ CHM
No
REPLACE
MAIN
AND
1M OHM
?
YES
atone
GO TO CHART 1
Trouble Isolation Chart for Attenuators
ASSEMBLY
GO TO CHART
1
5-9