Environmental
Conditions
AN
Power Requirements
Weight
Dimensions
Chapter 1: Genera! Information
General Characteristics
General Characteristics
These general characteristics apply to these oscilloscopes.
The instruments meet Hewlett-Packard's environmental specifications (section 750) for class
B-1 products with exceptions as described for temperature and condensation.
Contact your
local HP field engineer for complete details.
Temperature
Operating
0 °C to +55 °C (32 °F to +131 °F)
Nonoperating
—40 °C to +70 °C (—40 °F to +158 °F)
Humidity
Operating
up to 95% relative humidity (non-condensing) at +40 °C (+104 °F)
Nonoperating
up to 90% relative humidity at +65 °C (+149 °F)
Altitude
Operating
up to 4,600 meters (15,000 ft)
Nonoperating
up to 15,300 meters (50,000 ft)
Vibration
Operating Random vibration 5 to 500 Hz, 10 minutes per axis, 0.3 grms
Nonoperating
Random vibration 5 to 500 Hz, 10 minute per axis, 2.41 grms;
Resonant search, 5 to 500 Hz swept sine, 1 Octave/minute sweep rate, 0.75g, 5 minute
resonant dwell at 4 resonances per axis.
Voltage
115/230 Vac, -25% to +15%, 48 to 440 Hz.
Power
350 VA maximum
Net Approximately 11.8 kg (26 lb)
Shipping
Approximately 21.3 kg (47 Ib)
Refer to the outline drawings on the next page.
Notes
1
Dimensions are for general information only. If dimensions are required for building special
enclosures, contact your HP field engineer.