Main Assembly
To adjust the high-frequency pulse response
9 Connect the mains power cable and apply power.
Adjustment Procedure
Perform this procedure on the necessary channel only. The adjustment for each channel is
reached through a hole in the heatsink over the acquisition hybrid for that channel Use the
following illustration to locate the adjustments.
Pulse response
adjustment
sencod@sacaes
BoE%s
pee
a e s
:
©
54542216
Main Assembly Adjustment Locations
1 Connect the pulse generator output to the input of the channel needing adjusting.
2 Press the appropriate channel number key (bottom row) to select the channel menu.
3 Use the input resistance softkey (second from bottom) to select 502 be.
4 Press autoscale
, then set the following parameters.
Menu (hardkey)
Parameter
Setting
Vertical
(sensitivity)
40 mV/div
{channel key)
Horizontal
(time/div)
5 ns/div
(Setup)
(mode)
repetitive
Display
# of avg
16
(graticule)
grid
47