Refer to this procedure if the system is being moved or installed. Before moving the AP3600 system, contact the local Hitachi Chemical
Diagnostics service representative for detailed information.
WARNING:The AP3600 Instrument weighs approximately 170 kg (375 lbs.). Four persons are required to assist when lifting
the system. Refer to "Heavy Objects" on page 1-4 for details.

Installing a New System

1. Prepare the system location. The AP3600 instrument must be installed on a stable, flat surface that meets the following specifica-
tions (see Figure C-1 and Figure C-2):
Surface Dimensions: 1800 x 1060 x 1200 mm (W x H x D)
Supports a load of 185 kg
300 mm clearance (Left side)
300 mm clearance (Right side)
300 mm clearance (Rear)
FIGURE C-1. Instrument Clearances (all dimensions in mm)

Moving and Installing the System

OPTIGEN® AP3600™ AUTOMATED INSTRUMENT USER MANUAL
C-1
Moving and Installing the System