Navigator 500
Sodium analyzer

5 Calibration

This section describes how to calibrate the analyzer once it is
operational. Calibrations are initiated via the Cal prompt
displayed on Operator pages, or via the Calibration Level and
Advanced Level menus or as a Scheduled calibration.
Note.
Calibration menus can be accessed from the
Advanced level only – refer to Section 5.1 to access
the Calibrate level.
If calibration standard solution values need to be
changed from their default values, the values must be
set from the Advanced / Calibrate / Calibration Setup
menu only – refer to Section 5.1 to access the
Calibrate level.
Caution.
Do not calibrate the analyzer until the wet-section and
transmitter are installed and ready for operation – refer
to Section 4, page 19.
Allow at least 1 hour for the wet-section to stabilize
before running a calibration.
OI/ASO550–EN Rev. G

5.1 Accessing the Configuration level menus

The configuration level menus are used to configure the
wet-section(s) and set parameter values – see Fig. 7.1, page 42
for an overview of menus.
To enter Configuration Level menus:
1. Press the
key (below the
Navigator
Tag 1
2.93
21.7
21.7
Na
The Operator menus are displayed:
Navigator
Operator Pages
4.28
4.28
Data Views
Logs
Alarm Acknowledge
Manual Hold
21.7
21.7
Autoscroll
Media Card
Enter Configuration
2. Press the
key to select the Enter Configuration menu
and press the
key (below the
The Access Level page is displayed:
Access Level
Read Only
Calibrate
Advanced
Service
Back
3. Use the
key to scroll to the Advanced access level
and press the
key (below the Select prompt) to enter
the top level Configuration menus.
Use the
/
keys to scroll between top level menus
and press the
key (below the Select prompt on each
top level menu) to enter that level.
Note.
Refer to Section 7, page 42, for menu
descriptions.
Refer to Section 6.1, page 29, for details of menu
navigation and parameter selection / adjustment.
5 Calibration
icon).
10 : 34: 52
2012–04–15
ppb
°C
CAL
10 : 34: 52
2012–04–15
ppb
°C
icon).
Select
21