Navigator 500
Sodium analyzer

7.1.3 Device Setup

Menu
Device Setup
Select
Exit
Menu
Initial Setup
Instrument Tag
Temperature Units
Reset to Defaults
Security Setup
Calibrate Password
Advanced Password
Reset Passwords
Range Change Setup
Range Change
Range 1 Span
Range 2 Span
Range 3 Span
48
This level is used to access standard setup parameters.
Comment
A 16-character alphanumeric device (transmitter) identification tag
displayed in the title bar on all Operator pages.
Selects the temperature units displayed (°C or °F).
Resets the configuration to the factory default settings.
Security access levels are provided, each protected by a password of
up to 6 alphanumeric characters.
Provides access to Calibrate level only.
(Not factory-set.)
Provides access to all configuration parameters.
(Not factory-set.)
Clears all passwords.
Notes.
Displayed only if 1 single-stream wet-section is connected.
Range 3 must always be the largest range, Range 1 must always
be the lowest.
Range Changing is disabled automatically when the sensor type
is changed.
When Enabled, Analogue Output 2, Relays 3 and 4 are configured
automatically for range changing and are not available via the
Input/Output sections.
For correct operation, Range 1 Span should be the lowest range and
Range 3 Span should be the highest range.
When Enabled – the following zero and span ranges 1 to 3 are set
automatically:
Range
Zero
Span
1
0
100
2
0
1000
3
0
10000
The range spans can be changed to suit end-user's requirements.
7 Menu descriptions
Default
Navigator
°C
Disabled
Continued on next page...