Sodium analyzer
Icon
Diagnostic message
Media Near Full
No Low Cal
(S1, S2, S3, S4)
No High Cal
(S1, S2, S3, S4)
No Regen.Sol
(S1, S2, S3, S4)
No Sample
(S1, S2, S3, S4)
Table A.1 Diagnostic messages (Sheet 3 of 3)
74
Possible cause and suggested action
Memory card is more than 90% full.
Replace memory card.
Out of low calibration solution.
Replace low calibration solution and start a calibration.
. Displayed only if Solution Detection is enabled.
Note
Out of high calibration solution.
Replace high calibration solution and start a calibration.
. Displayed only if Solution Detection is enabled.
Note
Out of regeneration solution.
Replenish solution.
. Displayed only if regeneration hardware is fitted and solution detection is enabled.
Note
Sodium multi-stream only
No sample available at wet-section.
Check sample flow rates are >50 ml/min.
Appendix A – Troubleshooting