Preparing The Sodium And Reference Electrodes For Storage - ABB Navigator 500 Series Operating Instructions Manual

Sodium analyzer
Hide thumbs Also See for Navigator 500 Series:
Table of Contents

Advertisement

Navigator 500
Sodium analyzer
8.5.1 Preparing the sodium and reference electrodes for
storage
Referring to Fig. 8.4:
1. Unscrew (but do not remove) plastic sleeve
carefully slide sodium electrode
2. Fill the rubber teat
electrode) with High Cal solution or typically with 1 mg kg
sodium containing a few drops of concentrated ammonia
solution and push the teat over the end of the sodium
electrode.
3. Carefully withdraw reference electrode
flowcell.
4. Fill the rubber teat
electrode) with salt bridge solution and push the teat over
the end of the reference electrode.
5. Refit the filling hole plug
H
(reservoir option) to seal the refill aperture.
Note. Do not let electrodes dry out.
Fig. 8.4 Preparing the sodium and reference electrodes for
storage
64
A
B
out of the flowcell
D
(supplied with the sodium
E
from the
F
(supplied with the reference
G
(non-reservoir option) or
and
C
.
-1
8 Maintenance
OI/ASO550–EN Rev. G

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Navigator 550Navigator 540

Table of Contents