10. Before Calling a Service Engineer

10.3 Error

Error
Code
Insufficient tmp disk space
-2
Output the saved data or erase the backed up data.
Insufficient que disk space
-3
Output the saved data or erase the backed up data.
Sensor Unit: SelfCheck Error
-120
Detected an Error.
Sensor Unit: Sequence Error
-121
Sequence Error. Call a service engineer.
Sensor Unit: Control Error
-122
Detected Control Error. Call a service engineer.
Sensor Unit: PLD Error
-123
Detected PLD Error. Call a service engineer.
Sensor Unit: Grid Error
If you select [OK], the system will separate the Grid component, then change to
-124
force-Grid mode. In this case, the image quality cannot be guaranteed. If you
select [CANCEL], Grid component will be re-initialize.
Sensor Unit: Grid Error
-125
Because of the Grid error, this image may have a quality problem.
Sensor Unit: Temperature Error
-126
The internal temperature of the Sensor Unit exceeded the temperature limit.
Cannot capture images until the temperature goes below the limit.
Sensor Unit
Sensor Unit cannot become ready for exposure.
-127
Check the following part:
(message)
Sensor Unit: Signal Line Error
-128
Problem has occurred between sensor unit and EO box. Call a service
engineer.
Sensor Unit: Data Transmit Error
-129
Failed to transmit the data.
Capture Board Error
-130
Detected Capture Board Error.
Capture Board Control Error
-131
Detected Capture Board Control Error. Call a service engineer.
Error Message (Upper box: Error type
Lower box: Error description)
– 114 –
Remedy
Touch [OK] to close the window, and then see
6.4 Deleting the Image(s) (→ page 58), and
delete the image.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
• This cuts off control of the movable grid.
The grid lines appear over the image, and the
image quality cannot be guaranteed. To
continue with the exposure anyway, touch
[OK].
• Touch [CANCEL] to re-initialize the grid
unit.
Touch [OK] to close the window, and then
check the image. Perform exposure again if
there are any abnormalities.
Touch [OK] to close the window and try again
after a while.
Touch [OK] to close the window and lock the
slide adapter (optional unit) or the grid unit.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
Touch [OK] to close the window. As image
quality will not be guaranteed, take an image
again.
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.

Hide quick links: