Error
Code
Reject Info Output Error
-533
Selected directory for reject list can only be read. Reject info cannot be saved
on the file.
Reject Info Output Error
-535
Selected directory for reject list does not exist, or attribution is incorrect. Reject
info cannot be saved on the file.
Reject Info Output Error
-537
The file format for reject list includes incorrect strings. Reject info cannot be
saved on the file.
Reject Info Output Error
-539
Error has occurred on saving reject info on the file. Do you retry it?
DICOM Connect Error
-540
Transmission of the storage commitment request failed. Connection to output
equipment is not possible. Check the network or port number settings. Retry?
DICOM Transfer Error
-541
Transmission of the storage commitment request failed. An error occurred
during transmission with the output equipment. Retry?
DICOM Parameter Error
-542
Transmission of the storage commitment request failed. The DICOM output
parameter settings are invalid. Retry?
DICOM Response Error
Transmission of the storage commitment request failed. The DICOM response
-543
reported by the receiving equipment was an error. Check the receiving
equipment. Retry?
Storage Commitment Request Error
-548
The storage transfer was canceled, and so the storage commitment request
was not issued.
Calibration Error
-6000
(Sensor unit ID: 1 to 4): (Elapsed time) have passed since the last calibration.
Perform calibration.
Error Message (Upper box: Error type
Lower box: Error description)
10. Before Calling a Service Engineer
– 121 –
Remedy
Touch [OK] to close the window, and then see
1.2 Shutting Down the CXDI (→ page 5),
and turn off the CXDI.
When these errors occur, note the error code
and contact a Canon representative or
distributor.
• Touch [RETRY] to store the image on the
storage again.
• To cancel save, touch [CANCEL].
• After confirming the storage is normal,
touch [RETRY].
• Touch [CANCEL] to close the window and
then check the settings.
Touch [OK] to close the window.
Touch [CLOSE] to close the window, and then
see 8. Calibration (→ page 97) to perform the
calibration.