10. Before Calling a Service Engineer
Error
Code
Sensor Connection Error
-149
Multiple sensors sharing the same BodyPart cannot be used at the same time.
Turn on the power for one sensor only that will be used.
Calibration Error
-302
Too Much Dose. Or collimator is used.
Calibration Error
-303
Too Little Dose. Or collimator is used.
Calibration Error
-304
Invalid Data. Or collimator is used.
Calibration Error
-305
Collimator may be used.
QA Process Error
-306
Image Analysis Error. Adjust image process parameter manually.
Calibration Error
-307
Periodic noise is detected in acquisition data. It is necessary to remove the grid
before calibration.
REX Warning
REX is (detected value). It is below lower limit (setting values). Check generator
-310
setting and exposure conditions. If exposure conditions were correct, adjust IP
parameters.
REX Warning
REX is (detected value). It is over maximum limitation (setting values). Check
-311
generator setting and exposure conditions. If exposure conditions were correct,
adjust IP parameters.
Incomplete End Error
-402
The last image in the last study was not saved completely.
Incomplete End Error
-403
Do you want store captured images into internal storage in order to restore
them later?
External Storage Error
-509
Selected directory or disk for storage can only be read.
External Storage Error
-512
Cannot recognize the external storage. Set the disk properly and retry.
Media Control Error
-513
Operation has been canceled because the external storage or directory is busy.
Error Message (Upper box: Error type
Lower box: Error description)
– 116 –
Remedy
Touch [OK] to close the window. Then, turn off
the power of unused sensor unit(s).
Turn on only one sensor unit when multiple
sensor units share the same body part.
Touch [OK] to close the window. Adjust
exposure dose (mAs) or focal distance and take
an image again.
• Touch [OK] to close the window and repeat
the exposure.
• Please contact Canon if the irradiation field
aperture is not being used.
Touch [OK] to close the window, and then see
7. Processing the Image (QA Mode
Screen) (→ page 65) to manually perform
QA processing.
Check that the grid has been removed, and then
retake.
When calibration is to be performed for the
CXDI-40G COMPACT, ask the service
technician to do this job.
Touch [OK] to close the window and adjust the
dose.
Touch [OK] to close the window and repeat the
exposure.
• Touch [OK] to recover.
• Touch [CANCEL] to delete the image.
Touch [OK]. Change the attribute of the storage
medium and try again.
Touch [OK] to close the window and check the
setting of the external storage medium.
Touch [OK] to close the window and try again
after a while.