Err 18
CAL 19
* See section "Monitoring the kV- and mA-controls"
** Wiring diagram X037E
Tab. 1
Siemens-Elema AB
Solna, Sweden
Mode of
Cause
operation
Generator ON
The change in volt-
The kV value
age level in the
changes via the
capacitor bank takes
kV+ or kV- but-
too long
ton.
(>0.5 s/V). Refer
also to "Err 17".
Generator goes
The leakage current
ON before the
in the capacitor bank
ready lamp
is too large.
glows.
SPR8-215.898.02
Rev. 01
Possible reasons and advice
Check whether S10 on D7 is
switched off.
Check if any of the capacitors in the
capacitor bank are defective. The
output energy may be consumed
by a defective capacitor, refer to
section " Capacitor bank".
Check the signals
"CHARGE_HIGH", "DISCHARGE"
and "CHARGE_LOW", refer to
page 130**. Check fuse U1 on D7,
U5 and U8 on D8 and resistors
R100 and R101.
The measurement is performed dur-
ing charging at 150 V DC. Charging
is interrupted for 5 s.
If the voltage drops more than 1.6 V,
a message will be displayed.
See also "Err 17".
In cases where the generator has
not been operated for several
months, the leakage current in the
capacitors has increased. Use the
formation program P06, refer to
section "P06 - Capacitor
formation".
If the unit was in regular use before
the message occurred, the error
may result from increased leakage
current or a damaged capacitor,
refer to the section "Capacitor
bank".
Check the capacitance in the
capacitor bank, refer to the section
"P03 - Capacitor bank and
charging test". Refer also to page
130**.