Mechanical systems
Switch on the collimator lamp and check the compliance of the light and radiation fields
according to the specifications in the section on "Replacement of collimator".
Reattach all covers on the single tank.
Adjust the adjustment screw so that the single tank is balanced. If any problems occur,
refer to the section "Adjustment of arm system".
Use the service program "P16 - Show the exposure counter" and record the values of the
exposure counter in the service protocol and in the unit documentation.
Use the service program "P10 - Fast adaptation" to adjust the filament values to the new
single tank.
Reattach the remaining covers.
Siemens-Elema AB
Solna, Sweden
SPR8-215.898.02
Rev. 01
01.02
Page 17 of 28
MOBILETT Plus/ Plus E
SPS-UD
10 - 17