Mechanical systems
•
Switch on the collimator lamp and check the compliance of the light and radiation fields
according to the specifications in the section on "Replacement of collimator".
•
Reattach all covers on the single tank.
•
Adjust the adjustment screw so that the single tank is balanced. If any problems occur,
refer to the section "Adjustment of arm system".
•
Use the service program "P16 - Show the exposure counter" and record the values of the
exposure counter in the service protocol and in the unit documentation.
•
Use the service program "P10 - Fast adaptation" to adjust the filament values to the new
single tank.
•
Reattach the remaining covers.