Siemens MOBILETT Plus Service Instructions Manual page 59

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

Capacitor bank
Fig. 3
MOBILETT Plus
Fig. 4
MOBILETT Plus E
Measure the voltage between the + and - bars; the voltage must be less than 10 V.
Note the cable connections to the capacitor bank (Fig. 3 for MOBILETT Plus and Fig. 4
for MOBILETT Plus E)
Siemens-Elema AB
Solna, Sweden
- bar (neg)
- bar (neg)
SPR8-215.898.02
Rev. 01
01.02
+ bar (pos)
+ bar (pos)
Page 3 of 6