Mechanical systems
NOTICE
Set S10 to the OFF position.
Reinstall all covers except single tanks covers "A" and "B".
Make sure that the screw mountings between the fork and the arm are tightened
properly.
Remove the rope between the handle and the foot while keeping the single tank on the
table.
Tighten the adjustment screw until the single tank and arm are balanced. Refer to the
section "Adjustment of arm system".
Switch on the collimator lamp and check the compliance of the light and radiation fields
according to the specifications in the section "Replacement of collimator".
Reattach the remaining covers.
Siemens-Elema AB
Solna, Sweden
Do not stretch the cable too tightly in the arm.
There must be enough slack so that the cable is not pulled by the
mechanical movements.
SPR8-215.898.02
Rev. 01
Page 21 of 28
01.02
SPS-UD
10 - 21
MOBILETT Plus/ Plus E