Procedure to Access Calibration Constants

In the adjustment procedures, calibration constants stored in working memory will be changed
(adjusted). A calibration constant can be accessed by pushing the following key sequence:
[SHIFT] [ GHz] [number of the Cal. Constant, 1 to 99] [Hz]
[SHIFT][MHz]
[1]
[2] [Hz]
[SHIFT][kHz] [2] [2] [Hz]
The ENTRY DISPLAY will indicate the calibration
the right. Use the STEP keys to select a different calibration constant number. The ENTRY keyboard
or the front panel rotary knob can be used to change the value. After the new value has been entered,
press the [Hz] key to retain the new values in "working memory area.

Procedure to Store the Calibration Constants

After adjustments of the calibration constants are completed, the data thus stored in the "working
memory area" should be copied to the "protected memory area" by pressing the following key
sequence:
[SHIFT][MHz] [1] [4] [Hz]
[SHIFT][kHz][5] [3] [4] [9] [Hz]Wait
DISPLAY and then press [INSTR PRESET].
HP 8340B/41B
constant number on the left and the present value on
"CALIBRATION RESTORED" to be displayed in the ENTRY
for
Adjustments
5-3