Instrument
Signature Multimeter
Oscilloscope
Oscilloscope Probe
Digital Voltmeter (DVM)
Power Meter
Power Sensors
Frequency Counter
Universal Counter
Spectrum Analyzers
Frequency Standard
Synthesized Sweeper (used
as Local Oscillator in
~1
A 1, and Sweep Generator in
Al)
HP 8340B/41B
Table 4-2. Recommended Test Equipment (1 of5)
Critical Specifications
Clock Frequency: =20MHz
Dual Channel
Bandwidth: dc to 100 MHz
Vertical Sensitivity: =5 mV/Div
Horizontal Sensitivity:
50 ns/Div
10 Magnifier
External Sweep Capability
50 Ohm Vertical Input
10:1 Divider Probe
Range: —50 to +50 Vdc
Accuracy -'-001%
Input Impedance: =10 MOhms
Power Range: 1~aW to 100 mW
Accuracy: ±0.02dB
Frequency Range: 10 to 50 MHz
Power Range: 1 jiW to 100mW
Frequency Range: 50 MHz to maximum
Frequency of DUT
Power Range: 1
Frequency Range: Same as DUT
Input Impedance: 50 Ohms
Frequency Accuracy:
time Base Accuracy
Time Interval Range: 100 ns to 200 s
Frequency Range:
HP 8340B DUT: 0.01 to 22 GHz
HP 8341 B DUT: 0.01 to 20 GHz
Must have External Time Base Input
Center Frequency Accuracy in 0 Hz Span:
Same as Time Base Accuracy
Minimum Resolution Bandwidth: :=300Hz
Residual FM: <100 Hz
Log Fidelity: =1dB
Must have Video Output
Frequency Range: 20 Hz to 40 MHz
Resolution Bandwidth: =3Hz
Frequency: 10 MHz
Stability: >1 x 10 iD/yr
Frequency Range: Same as DUT
and
Must have External Time Base Input
Frequency Accuracy:
Same as Time Base Accuracy
RF Power Output: = +4 dBm
Phase Noise: Must meet or exceed HP
8340B/41 B
Line Related Spurious Signals:
Must meet or exceed HP 8340B/41 B
1iW to 100 mW
1 count
±
3
HP 5005A
HP 1741A
HP 10004D
P, A, T
HP 3455A or
HP 3456A
P, A,
T
HP 436A
Option 022
P, A, T
HP 8481A
A, T
HP 8485A
P,
P, A, T
HP 5343A
Option 011
P, A, T
HP 5316A
P, A, T
HP 8566A or
HP 8566B
P. T
HP 3585A
P, A
HP 5061A
P, A
HP 8340A
Opt. 001
4-3
Performance Tests