HP 8340B Manual page 167

Synthesized sweeper (including options 001, 004, 005,006, and 007)
Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

5-10. YO DELAY COMPENSATION, A54 (Cont'd)
YO DELAY ADJUSTMENT
Connect the test equipment as shown in Figure 5-29.
4.
5. Set the OUT as follows:
a. Press [INSTRPRESET]
b. Press [STARTFREQ] then enter [6]
Press [STOPFREQ] then enter [1] (3]
c.
d. Press [Ml] then enter [7]
e. Press [AMTDMKR] on.
Press [SAVE]then enter [2]
f.
g.
Press [SWEEPTIME] then enter [2] [0] [0] [msec]
h.
Press [SAVE]
Press Frequency Marker [OFF]
j.
Press [M2] then enter [1] [3]
Press [SAVE]then enter [3]
k.
pad, press [AUTO](sweep time)
I.
At ENTRY
(SAVE] then enter [4]
m. Press
n. Press [RECALL]then enter [1]
HP 8340B/41B
BOTTOM VIEW
Figure 5-30. YO Delay Compensation Adjustments Location
[.]
[.1
[2] [GHz]
[1]
then enter
[2] [GHz]
[.]
[9] [GHz]
[5] [GHz]
[.]
A54 YO
PR ETUN E
DAC/OLY COMP
Pw
A54R36
COFF
A54R32
CGN
A54R30
Adjustments
5-59