4-3. SWEEP TIME ACCURACY (Cont'd)
3.
Set up the 5316A as follows:
TIA—~B
GATE TIME
LEVEL/SENSE (CHANNEL A and B)
TRIGGER (CHANNEL A and B)
AC/DC (CHANNEL A and B)
ATTN (CHANNEL A and B)
FILTER
SEP/COM A
Channel A Triggering
Channel B Triggering
To ensure that an accurate sweep time indication is obtained from
the HP 5316A in step 4, allow the DUT to complete at least three
sweeps.
4. On the HP 8340B/41B, press [SWEEP TIME] [1] [0] [msec]. Verify that the sweep time
displayed by the HP 5316A is within the limits shown in Table 4-6.
5. Repeat step 4 for values shown in Table 4-6.
Selected
10 ms
lOOms
is
lOs
50 s
4-20
Performance Tests
NOTE
Table 4-6.
Sweep Time Accuracy Limits
Lower Limit
9.S ins
9Sms
.9Ss
9.Ss
47.S s
IN
MIN
+MAX
LEVEL
DC
Xl
NORM
COM A
Leading Edge
Trailing Edge
Upper Limit
10.S ms
lOSms
l.OSs
10.Ss
52.S s