Installation of biopsy controller holders

The biopsy controller is attached to the rear cover of the stand. To install the holders
proceed as follows:
1.
2.
3.
4.
Fig. 4
Rear cover, from the back side
Opdima?
Installation
Attach the two drilling templates (1/Fig. 4) included in this document on the inside
of the rear cover according to Fig. 4. The template marked B shall be placed
underneath the template marked A with no space between them. Make sure that
the templates are centered and that the sides of the templates are parallel to the
rear cover sides.
Drill five holes with a 5 mm drill in the rear cover according to the templates.
Install the three holders (2/Fig. 4) by using the nuts (3/Fig. 4).
Install the two standoffs (4/Fig. 4) by using the screws (5/Fig. 4) and the washers
(6/Fig. 4).