CAUTION
14.
It is advisable to use an extension cord with at least three outlets.
CAUTION
Opdima?
Installation
Ensure the line voltage selector switch (2/ Fig. 25) is set to the appropriate setting:
115V or 230V.
Plugging a 115V power cord into a 230V connector will severely
damage the system.
Connect the AC power cords of the main unit, the monitor, the MO unit, the local
Opdima? printer (if present) and the MAMMOMAT printer (if present) to the wall
socket marked with label.
The workstation and the printer must be grounded to the same
potential as the MAMMOMAT.
Connect the wall socket for the workstation and printer power
supply to the same branch circuit as the MAMMOMAT power sup-
ply.
(The workstation table can be connected to any branch circuit.)