6 - 2

Starting up the MAMMOMAT

Switch on the MAMMOMAT.

Starting up the workstation and login

The workstation is started according to instructions in Starting up the workstation and
login in Supplement to the Instructions for Use MAMMOMAT 3000 - Opdima?.
Use the user name "service" and the password obtained from Siemens service center
when entering the system.
To obtain optimal picture, adjust the monitor settings as follow:
Brightness: 5060
Contrast: 100

Starting up the MO unit

1.
2.

Installation of software

The software is installed in the factory before delivery. If the software for any reason
needs to be reinstalled, refer to Reinstallation of software in MAMMOMAT 1000/3000/
3000 Nova - Opdima? Service Instructions.

Mounting of covers

Mount the covers previously removed.

Filling in of LINA CARD

Fill in the LINA CARD and send it to the address stated on the card.

Performing of DHHS maintenance

In the USA, perform maintenance according to MAMMOMAT 300/3000, MAMMOMAT
3000 Modular incl. Stereotactic Biopsy Attachment and Opdima? DHHS Maintenance
Opdima?
Installation
HEWLETT
PACKARD
Fig. 1
Magneto-optical drive (HP)
Insert the MO disk, type 2.3 GB MO disk
For switch settings, SCSI ID and termination see Ultra 10/Ultra 1 MO unit in
MAMMOMAT 1000/3000/3000 Nova - Opdima? Service Instructions.
Register 3
SPB7-230.033.10
Rev. 03
08.02
Final procedures
Page 2 of 4
Siemens-Elema AB
SPS-UD
6
6
6
6
6
6
6
Solna, Sweden