Final procedures
General
The following procedures shall be performed:
•
Starting up the workstation, see Page 6 - 2
•
Country settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service
Instructions
•
Test of units,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service
Instructions
•
CCD camera calibration,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service
Instructions
•
Calibration of the biopsy unit,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service
Instructions
•
Verifying the calibration of the biopsy unit,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdima?
Service Instructions
•
Check of Opdima? AEC,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdima?
Service Instructions
•
Check of resolution,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdima?
Service Instructions
•
Activate DICOM (optional),
see DICOM option in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova -
Opdima? Service Instructions
•
Network settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdima? Service
Instructions
•
Printer setup (optional),
see Use of printer setup in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova
- Opdima? Service Instructions
•
Mounting of covers, see Page 6 - 2
•
Protective earth measurement,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdima?
Service Instructions
•
Filling in of LINA CARD, see Page 6 - 2
•
Performing of DHHS maintenance, see Page 6 - 2
NOTICE
Siemens-Elema AB
Solna, Sweden
The Opdima? external diaphragm must be used for tests and cali-
brations involving radiation.
Register 3
SPB7-230.033.10
Rev. 03
Page 1 of 4
08.02
SPS-UD