Monitoring SpO2
Masimo
11-10
Option
SAT SECONDS
RESPONSE
The following settings control pulse oximetry monitoring for the Masimo
configuration option.
Dash™ 3000/4000/5000
Function
Select the amount of time that SpO
outside the set limits before an alarm sounds.
OFF — Any alarm limit violation instantly alarms.
10 to 100 — Saturation seconds. Selectable in 5
saturation second intervals. The default setting is OFF.
The method of saturation seconds calculation is as follows:
The number of percentage points that the SpO
falls outside the alarm limit is multiplied by the number of
seconds that it remains outside the limit. This can be stated
as the equation "points x seconds = saturation seconds,"
where points equals SpO
percentage points at or outside
2
the limit, and seconds equals the number of seconds SpO
remains at that point outside the limit.
The upper and lower SpO
alarm limits are set in the same
2
way as traditional alarms.
NOTE
When Sat-Seconds is turned on, the signal strength
indicator is not displayed. In this event, check the
amplitude of the waveform to determine signal
strength. Be sure the SIZE (waveform scale) setting is
1X.
Select the averaging time (speed of response) to changes in
a patient's SpO
measurements.
2
Normal — Reports changes within four to six seconds
under motion-free conditions.
Fast — Reports changes in less than four seconds
under motion-free conditions.
saturation may be
2
saturation
2
2
2000966-386D