Cautions; Impact Or Results Of Improper Cleaning Products And Processes - GE Dash 3000 Operator's Manual

Patient monitor
Hide thumbs Also See for Dash 3000:
Table of Contents

Advertisement

Maintenance

Cautions

Impact or results of improper cleaning products and processes

A-4
4.
Dry thoroughly with a dry lint-free cloth and let air dry for at least 30 minutes.
NOTE
Drying times may vary based on the environmental conditions.
5.
Take care not to let fluid "pool" around connection pins. If this should happen,
blot dry with a soft lint-free cloth.
6.
Do not use excessive drying techniques, such as oven, forced heat, or sun drying.
Use the following precautions to prevent damaging the equipment:
Never immerse the device in any liquid.
Do not pour or spray any liquid directly on the device or permit fluid to seep into
connections or openings.
Never use conductive solutions, solutions that contain chlorides, wax, or wax
compounds to clean the device.
Never use solutions or products that contain the following:
Any type of Ammonium Chloride such as, but not limited to, Dimethyl
Benzyl Ammonium Chloride and Quaternary Ammonium Chloride
solutions.
Abrasive cleaners or solvents of any kind.
Acetone
Ketone
Betadine
Alcohol-based cleaning agents
Sodium salts
Never autoclave or steam clean the device.
Always use household glass cleaner to clean the touchscreen or standard
displays.
The following consequences can occur if you do not follow the cautionary guidelines
when cleaning the equipment:
Product discoloration.
Melting, dulling, or distorting of the device case.
Brittle and breaking device case.
Blurring the lettering on label text.
Equipment malfunction or failure.
Void warranty.
Dash™ 3000/4000/5000