Canon CXDI-40G COMPACT Service Manual page 120

Table of Contents

Advertisement

3-4) Changing the items below contained in the file allows you to change the settings for
the counter in the user screen.
ExpResult.ini file
[SystemCounter]
TotalStudy=
UsrStudyCounter=
UsrExposure
Counter=
[LastCounterReset]
StudyCounter=
ExpCounter=
[COUNTER 0]
TotalCount=
[COUNTER1]
TotalCount=
[COUNTER2]
TotalCount=
[COUNTER3]
TotalCount=
3-5) After overwriting the values, overwrite the file "ExpResult.ini" and save.
3-6) Close all windows that are open on the Desktop, and then start up the CXDI
application.
3-7) Open the "System Information" screen, and check that the changed items have been
set correctly. [Fig 2]
"TOTAL IMAGES" indicates the total number of images obtained with all sensors
automatically.
2
.
I
n
s
t
a
l
l
a
2
.
I
n
s
t
a
l
l
a
Screen Display
TOTAL STUDIES
STUDY COUNTER
IMAGE COUNTER
Year, month, day, hour, minute
Year, month, day, hour, minute
Total number of images obtained
with sensor unit 1.
Total number of images obtained
with sensor unit 2.
Total number of images obtained
with sensor unit 3.
Total number of images obtained
with sensor unit 4.
[Fig.2]
- 100 -
t
i
o
n
t
i
o
n
Can be overwritten at the
user screen
Can be overwritten at the
user screen
Note