(3) Cable connections
1) Loosen the SKINTOP spirals and lock nuts for the sensor cable and X-ray I/F cable,
mount the cables to the power box, and then use a wrench to secure the SKINTOP spirals
and lock nuts. Be sure that the cables have enough length to connect to the power box
connectors.
Note: To prevent damage in case the cable is pulled by a strong force, adjust by tightening a
SKINTOP spiral as shown in the figure below.
<Sensor cable>
2) Attach the cable clamp to the sensor cable, and then fix it to the power box using the screw
(M4
6 mm).
×
Screw (M4×6 mm)
2. Installation
Sensor cable SKINTOP spiral
Gap 0.5 to 1.0 mm
Clamp the cable's shield portion
- 20 -
X-ray I/F cable SKINTOP spiral
Lock nut
No gap
<X-ray I/F cable>