< I n s p e c t i o n
ITEM
Category
1
Imaging unit
2
Power box
Control PC
3
System connections
4
Calibration
5
6
Grid installation
7
Self-diagnosis
NO
Item
1
Sensor cable
Sensor cable
2
connector
3
Ready Lamp
Ready Lamp
4
Connector
5
Firmware version
6
PLD code version
Imaging unit IP
7
address
1
Connector
2
Power switch
1
Connector
1
Startup
2
CCR startup
3
Status lamps
4
Shutdown
X-ray generator
1
settings
Preparation for
2
exposure
Calibration
3
processing
Grid installation
1
angle check
1
X-ray generator
CXDI-40G COMPACT Inspection Check Sheet
I t e m s >
Characteristics values
The cable must not be
flattened or twisted and
its covering must not be
damaged.
The connector must be
connected securely,
and there must be no
play.
The ready lamp must
not be flattened or
twisted and its covering
must not be damaged.
The connector must be
connected securely.
-
-
-
The connector must be
connected securely.
It must be possible to
set this switch to its ON
and OFF positions
properly.
The connector must be
connected securely.
The system power must
come on when the
power of the control PC
is turned on..
The system must start
up with no errors.
The lamps must provide
the correct indication
that corresponds to the
sensor statuses.
The system power must
go off when the power
of the control PC is
turned off.
The conditions under
which calibration is
implemented must be
noted..
Phototimer must be
installed. Grid must not
be installed.
"Calibration completed
successfully."
Within ±1.0 degree
The conditions under
Page 1 of 2
Test means
Record
Visual check
Visual check,
touch
Visual check
Visual check,
touch
Visual check
Ver
Visual check
Ver
Visual check
Visual check,
touch
Visual check,
touch
Visual check
Visual check
Visual check
Visual check
Visual check
Audit_grid.exe
Visual check
kV