2. Installation
2) Look in the X-ray tube to check the cross line reflected from the mirror matches the cross line on
the X-ray tube. If it does not match, move it right and left or put the spacer before and behind the
stand base to align the slight differences.
Case the cross line
Look in
Reflect the cross line
Reflected cross line
Cross line of X-ray tube
Correct Alignment
Shifted Horizontally
Shifted Vertically
(4) Regardless the height of the X-ray tube and the sensor unit, it should keep the positional relation
aligned in step (3) above.
Center of X-ray tube and the sensor unit
should match in any height.
x
Should be perpendicular in
any height