Operation With Different Analyzer Modules - Siemens 7MB3000 Series Operating Instructions Manual

Wall-mounted device
Table of Contents

Advertisement

Combined operation
5.4 Application planning
5.4.6

Operation with different analyzer modules

OXYMAT 7/ ULTRAMAT 7/ CALOMAT 7
Table 5-4
Operating
mode
Pressure op‐
eration
Suction oper‐
ation
50
Mounting conditions and positions for the sample gas restrictor
Sample gas restrictor/clamping screw
OXYMAT 7/ ULTRAMAT 7
 
OXYMAT 7 must always be
flowed first.
Install a sample gas restric‐
tor in front of the sample
gas inlet of the OXYMAT 7.
ULTRAMAT 7 must always
be flowed first.
Install a sample gas restric‐
tor in the sample gas outlet
of the OXYMAT 7.
Replace a sample gas re‐
strictor in the sample gas
inlet of the OXYMAT 7 with
a clamping screw.
ULTRAMAT 7/ CALOMAT 7
We recommend flowing the
CALOMAT 7 as the first mod‐
ule
 
Operating Instructions, 07/2023, A5E31930403-AB
OXYMAT 7/ CALOMAT 7
No restrictions