Siemens 7MB3000 Series Operating Instructions Manual page 69

Wall-mounted device
Table of Contents

Advertisement

3. Remove the equipotential bonding cable
Note: The cable for equipotential bonding
module
Figure 7-3
4. Loosen the eight Torx screws
5. Check the flat seal
Replace the flat seal
6. Swivel the analyzer module
7. Secure the analyzer module
– Use eight Torx screws on the base of the housing
– Use two Torx screws on the rear panel with a torque of 2.5 Nm.
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
is installed.
Unlatching plug-in connectors at the cable equipotential bonding
NOTICE
Damage to the device through pulling the latched cable
Cables for equipotential bonding are equipped at their two ends with latching plug-in
connectors. Pulling on a latched cable can cause the plug-in connector to be torn off or its
contacts to be damaged.
• Before pulling out the plug connector, loosen the latch by pressing with your index finger.
for damage.
Installing / removing and connecting analyzer and option modules
7.4 Installing analyzer modules, standard version
from the blank plate of the respective slot.
is no longer required after the analyzer
and remove the blank plate at the corresponding slot.
if it is damaged or porous.
forward and insert it into the wall housing
in the slot as follows:
with a torque of 0.8 Nm.
.
69