Siemens 7MB3000 Series Operating Instructions Manual page 77

Wall-mounted device
Table of Contents

Advertisement

3. Remove the following cables:
– Power cable to the analyzer module printed circuit board and power supply unit
– 12-pin ribbon cable (CAN) to the analyzer module printed circuit board and the
– If applicable, 26-pin ribbon cable to the analyzer module printed circuit board or the
4. Loosen the eight Torx screws on the base of the housing
rear wall.
5. Swivel the analyzer module
6. If you are not installing and connecting a new analyzer module:
Fasten the blank plate over the respective slot using the eight Torx screws
of 0.8 Nm.
7. Close the door of the housing with the six screws
Tighten the screws
Wall-mounted device
Operating Instructions, 07/2023, A5E31930403-AB
Installing / removing and connecting analyzer and option modules
processing unit
analog output adapter card and the option module OM 2.1 or OM 2.2
on the door with a torque of 3.5 Nm.
7.8 Removing analyzer modules, standard version
forward and remove it from the wall housing.
and the two Torx screws on the
.
with a torque
77