Alignment Procedure; Introduction; Test Equipment Required; Preparation For Alignment - Motorola R-20010 Maintenance Manual

Communications system analyzer
Table of Contents

Advertisement

3.
1 INTRODUCTION
This section
provides
a basic (paragraph
3.2)
and an
extended (paragraph
3.3)
alignment
procedure.
The
basic procedure, which should accompany any service
work,
requires
only a calibrated oscilloscope.
The
extended procedure, which should be
performed
at
nominal six-month
intervals,
requires
the oscillo-
scope,
module
extenders, a calibrated
digital
voltme-
ter and other equipment as listed in Table 3-2. All
adjustments
not covered in
this
procedure
should
be
·
performed on suitable module test fixtures only.
3.1.1
TEST
EQUIPMENT
REQUIRED
The basic procedure
requires the
test
equipment
listed in
Table 3-1. The
additional equipment
required
for the extended
procedure is listed
in
Table
3-2.
Table
3-1. Basic
Test Equipment Required
Desai ption
Model
'Oscilloscope
Tektronix
2215
Test-Point
Shorting
Jumper
Nonmetallic
Alignment Tool
An R2001
is
a
suitable
substitute.
Table 3-2.
Extended
Test
Equipment
Required
Description
Model
OsciUoscope
Tek 2215
'Digital
Voltmeter
FlukeSMOA
RF Signal Generator
Fluke6060B
Modulation
Meter
HP8901
Audio Generator
HP8903
Receiver Test
Cover
Motorola
Extender
Card Set
Motorola
RPX-
4379A
• An
R2001
is a suitable substitute
for these
separ:~te
equipments.
3.1.2
PREPARATION
FOR
ALIGNMENT
1.
Perform all alignments at
normal
ambient
tem-
perature.
2. Remove the
top
cover of the unit to
be
aligned.
3.
Turn
on the unit to be aligned, allowing a warmup
time
of 15
minutes before
starting
alignment.
3-1
SECTION 3.
ALIGNMENT PROCEDURE
3.2
BASIC ALIGNMENT
PROCEDURE
3.2.1 LOW VOLT AGE
SUPPLY
1.
Remove low
voltage
power
supply
cover and turn
on
System
Analyzer.
2.
Connect a DVM
to
test point TP5 on the Output
board
A5.
3.
Adjust
R6
on
the
Control
board,
A4,
(Figure 3-1)
for a DVM
reading
a
TP5
(A5)
of +5.2
Vdc ±0.1
Vdc.
3.2.2
CRT INTENSITY BIAS
1.
Select
Scope DC display and Ext
Horiz
input
mode
on the
System Analyzer. Set the
Intensity
control
fully counter clockwise.
CAUTION
Do not
let
a dot
stay
in one
place
on
the
CRT
screen
for more than
30
seconds because
it
will
leave
a
permanent
burn
in
the
phosphor.
2. Adjust the
Intensity Bias
potentiometer on
t
he
Scope
Amplifier board
(Figure
3-2) until
a dot
appears on the
screen.
You may
have
to
use
the
Vertical
and Horizontal
position control
(Vert,
Horiz)
on the
front
panel to
bring
the dot
onto
t
he
screen.
Then
back off
t
he
Intensity Bias
poten-
tiometer until the dot just
disappears.
3.2.3 CRT
INTENSITY BALANCE
1.
Select Scope DC display and
a
horizontal sweep
rate of
1
mSec/Div
on
the System Analyzer. Set
the
horizontal time base vernier to calibrate
(Cal)
and
adjust the
Intensity
control
for
a barely
visible
horizontal line
on
the
CRT.
2.
Adjust
the
Intensity
Balance potentiometer
(Figure
3-2) for
uniform
intensity
of
t
he
horizon-
tal trace from left
to
right.
The Balance
poten-
tiometer
affects the
intensity
on
the left
side of
the
trace.

Advertisement

Table of Contents
loading

Table of Contents