HP 5344S Operating Instructions Manual page 70

Microwave source synchronizer
Hide thumbs Also See for 5344S:
Table of Contents

Advertisement

Modei 5344S
Operation and Programming
Description:
The 5344S locks the signal source to the specified CW frequency. The signal source is automatically tuned
by the 5344S. No manual tuning is required.
Connections:
1. The address of the 8350A Sweep Oscillator must be set to the same address as the 5344A. The top switch of
the 5344A ADDRESS switches must be set to the SYS CONT position.
HP-IB
Example:
CW locked frequency at 2.000000000 GHz.
1.
Set the 5344A HP-IB address to 00 by setting the bottom five switches to the right.
2.
Set the 5344A to the System Controller mode by setting the top HP-IB switch to the left (SYS CONT.)
3.
Set the 8350A address to 00 (equal to the 5344A) by pressing the following key sequence:
SHIFT
.
J
(blue key)
r
-
LCL
f
-'
0
...
GHz
The HP-IB address will be shown on the Frequency/Time display of the 8350A.
4.
Set MANUAL LOCK and AUTO LOCK off (lamp off) and verify the 5344A front panel CONT lamp is
on. Also verify the plug-in's CW Filter is off.
5.
Using the 5344A Function keys, press and release MODE until the CW annunciator lights. The
MODE key will scroll through the four modes of operation. If you pass CW, continue pressing
MODE until you return to CW.
Figure 3-12.
CW
in AUTO LOCK
3-40

Advertisement

Table of Contents
loading

Table of Contents