HP 5344S Operating Instructions Manual page 78

Microwave source synchronizer
Hide thumbs Also See for 5344S:
Table of Contents

Advertisement

Model 5344S
Operation and Programming
With the optional method:
8350A CNTR TRIG to the 5343A SWP INTFC A
After setting up the Lock/Roll sweep, you can count the markers or Stop frequency by:
1.
Pressing STOP on the 8350A. This selects which parameter is to be measured by the 5343A.
2.
Pressing:
(-
SHIFT
(blue key)
on the 8350A.
This enables the counter interface function on the 8350A.
Whenever the 5344S Option 043 is locked in Lock/Roll, the sweeper interface function of the 5344S is
automatically enabled. If the 5344S is not locked, simply press SWP M on the 5343A to enable the sweeper
interface.
If you wish to count one of the markers on the 8350A, the operation is quite similar exceptthat you would
select the marker you wish to count in Step 1 above, rather than the sweeper STOP frequency.
To disable the sweeper/counter interface feature, press:
SHIFT
(blue key)
n
on the 8350A.
Figure 3-15. Lock and Roll in AUTO LOCK (Continued)