Toshiba famio XG SSA-530A Operation Manual page 104

Diagnostic ultrasound system
Table of Contents

Advertisement

Guidance and manufacturer's declaration - electromagnetic immunity
The SSA-530A is intended for use in the electromagnetic environment specified below. The
customer or the user of the SSA-530A should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Electrostatic
±6 kV contact
discharge (ESD)
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
transient/burst
supply lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential
IEC 61000-4-5
mode
±2 kV common mode
Voltage dips,
<5 % UT
short interruptions
(>95 % dip in UT)
and voltage
for 0,5 cycle
variations on
40 % UT
power supply
(60 % dip in UT)
input lines
for 5 cycles
IEC 61000-4-11
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE 1: UT is the a.c. mains voltage prior to application of the test level.
NOTE 2: Do not use cables or parts other than those intended for this product.
If cables or parts other than those intended for this product are used, performance
related to electromagnetic immunity may deteriorate.
Compliance level
±6 kV contact
±8 kV air
±2 kV for power
supply lines
±1 kV for input/output
lines
±1 kV differential
mode
±2 kV common mode
<5 % UT
(>95 % dip in UT)
for 0.5 cycle
40 % UT
(60 % dip in UT)
for 5 cycles
70 % UT
(30 % dip in UT)
for 25 cycles
<5 % UT
(>95 % dip in UT)
for 5 s
3 A/m
No. 2B730-973EN*D
26-2
Electromagnetic
environment - guidance
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, the relative
humidity should be at least
30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the
user of the SSA-530A requires
continued operation during
power mains interruptions, it is
recommended for the
SSA-530A to be powered from
an uninterruptible power supply.
Power frequency magnetic
fields should be at levels
characteristic of a typical
location in a typical commercial
or hospital environment.