21.2 Maintenance Checks to Be Carried Out by Service
Personnel
The following checks are required to ensure the performance and safety of the system.
Contact your TOSHIBA representative when carrying out these checks, because they
require special techniques.
Check category
Cleaning
Electric safety
Mechanical safety
Image recording
21.3 Consumable Parts and Parts Requiring Periodic
Replacement
This system contains some parts requiring periodic replacement and some consumable
parts.
The consumable parts include the casters, fuses, monitor, lamp at the bottom of the
CRT monitor, etc. For replacement, special techniques are required. Contact your
TOSHIBA representative.
Check item
Filter
Interior of the system
Peripheral units
Protective conductor resistance
Ground line leakage current
Enclosure leakage current
Patient leakage current I
Patient leakage current III
Patient auxiliary current
Check of the casters
Check of the caster mounting sections
Check of the monitor mounting mechanism
Operating panel
Mounting mechanism for the peripheral devices
Other mechanical parts
External appearance of the transducer
Images in each mode
Image recording using the standard transducer
No. 2B730-973EN*D
21-5
Check interval
1 year
1 year
1 year
1 year
*