HP 3325B Remote Operation Command Set
SS; Start Single Sweep Command
The effect of the SS command depends on the state of the instrument. If the instrument is
not sweeping and not in the sweep-reset state, then the SS command puts the instrument in
the sweep-reset state at the sweep Start Frequency. If the instrument is already in the
sweep-reset state, this command starts a single sweep. If the instrument is sweeping, this
command stops the sweep and does not restart it.
Single sweeps can be started using the HP-IB Group Execute Trigger command. Before using
the GET command, the HP 3325B must be in the enhancements mode and the sweep must
be reset using the RSW command.
Command Availability
SS
" HP 3325B
Ye^
~
HP 3325A
Yes