HP 3325B Remote Operation Command Set
Table 2-51. Error Messages
Code
Description
FAIL 010
Hardware failure, DAC range
FAIL Oil
Bad checksum, low byte of ROM
FAIL 012
Bad checksum, high byte of ROM
FAIL 013
Machine data bus line stuck low
FAIL 014
Keyboard shift register test failed
FAIL 021
Signal too big during calibration
FAIL 022
Signal too small during calibration
FAIL 023
DC offset too positive during cal
FAIL 024
DC offset too negative during cal
FAIL 025
Unstable/ noisy calibration
FAIL 026
Calibration factor out of range: AG gain offset
FAIL 027
Calibration factor out of range: AC gain slope
FAIL 028
Calibration factor out of range: DC offset
FAIL 029
Calibration factor out of range: DC slope
FAIL 030
External ref unlocked
FAIL 031
Oscillator unlocked. VCO voltage too low
FAIL 032
Oscillator unlocked, VCO voltage too high
FAIL 033
HP-IB isolation circuits test failed self test
FAIL 034
HP-IB iC failed self test
FAIL 035
RS-232 test failed loop-back test
FAIL 036
,
Memory lost (battery dead)
FAIL 037
Unexpected interrupt
FAIL 038
Sweep-limit-flag signal failed seif test
FAIL 039
Fractional-N IC failed self test
FAIL 040
Modulation Source failed self test
FAIL 041
Function-integrity-fiag flip-flop always set
Error 100
Entry parameter out of bounds
Error 200
invalid units suffix for entry
Error 201
Invalid units suffix with high voltage
Error 300
Frequency too large for function
Error 400
Sweep time too large (same as sweep rate too small)
Error 401
Sweep time too small
Error 500
Ampiitude/offset incompatible
Error 501
Offset too big for amplitude
Error 502
Amplitude too big for offset
Error 503
Amplitude too small